Standards Download Free
BS pdf,ASME pdf,ISO pdf,ASTM pdf,AS pdf,GB pdf
ISO 9873-2019 pdf free
HomeISO StandardsISO 9873-2019 pdf free
DESCRIPTION

ISO 9873-2019 pdf free.Dentistry一Intra-oral mirrors.
The mirror casing/stem joint shall withstand the test load without any sign of weld failure. Distortion of the casing is not considered as a failure. Test in accordance with £5.
4.5 Handle
All external surfaces of the handle shall be visibly free from imperfections.
Test in accordance with Si.
4.6 Resistance to reprocessing
lntra-oral mirrors, including the handle, shall withstand 100 reprocessing cycles, as defined by the manufacturer’s instructions in accordance with ISO 17664, without deformation or showing signs of corrosion. There shall be no visible damage to the reflective coating and there shall be no ingress of water.
The manufacturer shall specify if the mirror handle shall be disassembled from the head for the reprocessing cycle.
The manufacturer shall specify the maximum number of reprocessing cycles for both the mirror and the handle if this is less than 100 cycles. In this case this number shall be used for the test, in place of the 100 cycles.
Test in accordance with £2.
5 Measurement and test methods
5.1 Visual inspection
Conduct visual inspection at normal visual acuity without magnification.
5.2 Resistance to reprocessing
Submit the mouth mirror to 100 reprocessing cycles according to the manufacturer’s instructions in accordance with ISO 17664. The reprocessing cycle shall include the manufacturer’s recommended methods of cleaning, disinfection and sterilization, in accordance with ISO 17664.
Inspect all surfaces of the instrument for any signs of corrosion or surface defects and inspect the reflective surface of the mirror for any loss of integrity.
NOTE Discoloratlons due to water stains are not signs of corrosion.
lithe manufacturer specifies a maximum number of reprocessing cycles allowed, this specified number shall he used in place of the 100 cycles.
The hole, key 1, which is 10 mm in diameter, has cross-wires positioned as shown and a light source situated immediately behind it. A white card, key 2, marked with squares as shown, is fixed close to the hole, so that the image of the cross-wires will be in the same horizontal plane as the cross-wires.
5.3.2 Procedure
Place the mirror (with or without the handle) as shown in Figure 4 and move it forward or backward until the sharpest image is obtained in the centre square of the card. Measure the focal distance, f, to 1 mm, between the mirror and the card. Calculate the nominal magnification, M, in accordance with 4.3.3. ISO 9873 pdf download.


Tags:
Related Downloads
  • ISO 8600-1-2013 pdf free download.Endoscopes  – Medical endoscopes and endotherapy devices – Part 1: General requirements. 6 Marking 6.1 Minimum marking Iach individual endoscope or endotherapy device shall have the lollowing minimum marking: a) model number and/or other mark sufficient to identify the endoscope or endotherapy device and its manufacturer; b) maximum insertion portion width, minimum instrument channel width, working length, field of view and/or direction of view where such identification is necessary for the intended use of the endoscope or endotherapy device. The Insertion portion width and instrument channel width units shall be in millimetres. The insertion portion width and instrument channel width can also be marked in French size as defined in 3.5, shown by either ‘Fr or an encircled number; c) wherever reasonable and practicable. the endoscope or endotherapy device and detachable component(s) shall be identified in terms of lot numbers or serial numbers, etc. 6.2 Marking legibility The marking shall remain legible over the lifetime of the device when the endoscope or endotherapy device is used, cleaned, disinlected, sterilized and stored in accordance with the instruction manual. 6.3 Marking exceptions When marking on the endoscope or endotherapy device or detachable component(s) is impossible to achieve due…

  • ISO 15688-2012 pdf free download.Road construction and maintenance equipment – Soil stabilizers 一 Terminology and commercial specifications. This International Standard establishes the terminology, definitions of operation and commercial specifications for soil stabilizers and their components intended for use in road construction and pavement works. It does not apply to soil stabilizers used in agricultural applications. 2 Normative references The following documents, in whole or in part, are normatively referenced in this document and are indispensable for its application. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.. ISO 3911, Wheels and rims for pneumatic tyres — Vocabulary, designation and marking ISO 6746-1, Earth-moving machinery — Definitions of dimensions and codes — Part 1: Base machine ISO 6746-2, Earth-moving machinery — Definitions of dimensions and codes — Part 2: Equipment and attachments ISO 7134. Earth-moving machinery — Graders — Terminology and commercial specifications 3 Terms and definitions For the purposes of this document, the following terms and definitions apply. 3.1 soil stabilizer self-propelled machine, either towed or transported, with the function of pulverizing, breaking-up, aerating, homogenizing, and loosening existing and imported soil or paving materials and mixing them…

  • ISO 18436-1-2012 pdf free.Condition monitoring and diagnostics of machines – Requirements for qualification and assessment of personnel一 Part 1: Requirements for assessment bodies and the assessment process. 5 Requirements for assessment body personnel 5.1 General provisions In order to ensure that the assessment process is carried out effectively and uniformly, the competence requirements for personnel involved in the entire process shall be defined by the assessment body and, in the case of a third party, be approved by the responsible TCC (in accordance with ISO/IEC 17024). The assessment body shall require its personnel (internal or external) to sign a contract, or other document, by which they commit themselves to comply with the rules defined by the assessment body, including those relating to confidentiality and those relating to independence from commercial and other interests, and from any prior or present link with the persons to be examined that would, in the opinion of the interested parties, compromise impartiality. Clearly documented instructions shall be available to the personnel, describing their duties and responsibilities. These instructions shall be maintained up to date. All personnel involved in any aspect of assessment activities shall possess appropriate educational qualifications, experience and technical expertise, which satisfy defined…