DESCRIPTION

ISO 7176-8-2014 pdf free.Wheelchairs – Part 8: Requirements and test methods for static, impact and fatigue strengths.
When tested in accordance with Clauses 8. 2., and 1, a single wheelchair shall meet all the following requirements at the conclusion of the tests.
a) No component shall show evidence of visible cracks, be fractured, or have become detached, with the following exceptions.
1) Readjustment of postural supports Is allowed after each of the tests in Clauses 8 and 9.
2) Re-tightening, readjusting, or refitting of components that are identified in the operator’s manual as operator-adjustable components is allowed at 25 % intervals during each of the multi drum, drop, and manual brake fatigue testing procedures (Clause 10). Operator adjustable components, as identified in the operator’s manual, may not be adjusted using tools unless the tools are supplied with the wheelchair. If there are operator adjustable components, fatigue test equipment may be stopped at 25 % plus or minus S % intervals, to determine if re-tightening, readjusting, or refitting of operator-adjustable components is required. Re- tightening, readjusting, or refitting shall then be performed, following the procedures outlined in the operator’s manual. Restart the test equipment after re-tightening, readjusting, or refitting has been performed.
3) Re-tightening, readjusting, or refitting of any other component is not allowed.
4) During the fatigue testing (Clause 10), the following wear items, if they are identified in the operator’s manual, may be replaced no more than twice per item: tyres (including solid tyres), inner tubes, drive belts, castor wheel rubber. In the case of castor wheels that are a single integral part, replacement is only permitted because of wear or failure of the running surface, but not for failure of the wheel structure or other elements (e.g. bearings). No other wear items may be replaced.
5) Cracks in surface finishes, such as paint, that do not extend into the structural material do not constitute a failure.
b) No externally visible electrical cable shall be cut, abraded, or crushed. No externally visible electrical connector shall be crushed or disconnected.
c) All parts intended to move, rotate or be removable, folding or adjustable shall operate as described by the manufacturer.
d) All power-operated systems shall operate as described by the manufacturer.
e) Handgrips shall not be displaced.
f) No component or assembly of parts shall exhibit visible plastic deformation, free play, or loss of
adjustment that adversely affects the function of the wheelchair.
g) The brake mechanism shall not have moved from the pre-set positions.
4.2 Disclosure requirements
Manufacturers shall disclose in their specification sheets, in the manner and sequence specified in
ISO 7176-15, ii the wheelchair meets the strength requirements of this part of ISO 7176.
5 Test apparatus
5.1 Loading device, shall be capable of applying forces (compressive or tensile) to the wheelchair in the range 15 N to 2 000 N to an accuracy of ±3 %.
5.2 Concave loading pad, shall be made of metal or hard wood as shown in Figure 1. If suitable, the concave loading pad specified in ISO 16840-3 may be used in place of the concave loading pad shown in Figure 1.ISO 7176-8 pdf download.


Tags:
Related Downloads
  • ISO 8600-1-2013 pdf free download.Endoscopes  – Medical endoscopes and endotherapy devices – Part 1: General requirements. 6 Marking 6.1 Minimum marking Iach individual endoscope or endotherapy device shall have the lollowing minimum marking: a) model number and/or other mark sufficient to identify the endoscope or endotherapy device and its manufacturer; b) maximum insertion portion width, minimum instrument channel width, working length, field of view and/or direction of view where such identification is necessary for the intended use of the endoscope or endotherapy device. The Insertion portion width and instrument channel width units shall be in millimetres. The insertion portion width and instrument channel width can also be marked in French size as defined in 3.5, shown by either ‘Fr or an encircled number; c) wherever reasonable and practicable. the endoscope or endotherapy device and detachable component(s) shall be identified in terms of lot numbers or serial numbers, etc. 6.2 Marking legibility The marking shall remain legible over the lifetime of the device when the endoscope or endotherapy device is used, cleaned, disinlected, sterilized and stored in accordance with the instruction manual. 6.3 Marking exceptions When marking on the endoscope or endotherapy device or detachable component(s) is impossible to achieve due…

  • ISO 15688-2012 pdf free download.Road construction and maintenance equipment – Soil stabilizers 一 Terminology and commercial specifications. This International Standard establishes the terminology, definitions of operation and commercial specifications for soil stabilizers and their components intended for use in road construction and pavement works. It does not apply to soil stabilizers used in agricultural applications. 2 Normative references The following documents, in whole or in part, are normatively referenced in this document and are indispensable for its application. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.. ISO 3911, Wheels and rims for pneumatic tyres — Vocabulary, designation and marking ISO 6746-1, Earth-moving machinery — Definitions of dimensions and codes — Part 1: Base machine ISO 6746-2, Earth-moving machinery — Definitions of dimensions and codes — Part 2: Equipment and attachments ISO 7134. Earth-moving machinery — Graders — Terminology and commercial specifications 3 Terms and definitions For the purposes of this document, the following terms and definitions apply. 3.1 soil stabilizer self-propelled machine, either towed or transported, with the function of pulverizing, breaking-up, aerating, homogenizing, and loosening existing and imported soil or paving materials and mixing them…

  • ISO 18436-1-2012 pdf free.Condition monitoring and diagnostics of machines – Requirements for qualification and assessment of personnel一 Part 1: Requirements for assessment bodies and the assessment process. 5 Requirements for assessment body personnel 5.1 General provisions In order to ensure that the assessment process is carried out effectively and uniformly, the competence requirements for personnel involved in the entire process shall be defined by the assessment body and, in the case of a third party, be approved by the responsible TCC (in accordance with ISO/IEC 17024). The assessment body shall require its personnel (internal or external) to sign a contract, or other document, by which they commit themselves to comply with the rules defined by the assessment body, including those relating to confidentiality and those relating to independence from commercial and other interests, and from any prior or present link with the persons to be examined that would, in the opinion of the interested parties, compromise impartiality. Clearly documented instructions shall be available to the personnel, describing their duties and responsibilities. These instructions shall be maintained up to date. All personnel involved in any aspect of assessment activities shall possess appropriate educational qualifications, experience and technical expertise, which satisfy defined…