Standards Download Free
BS pdf,ASME pdf,ISO pdf,ASTM pdf,AS pdf,GB pdf
BS EN ISO 8536-4:2020 pdf free
HomeISO StandardsBS EN ISO 8536-4:2020 pdf free
DESCRIPTION

BS EN ISO 8536-4:2020 pdf free.Infusion equipment for medical use Part 4: Infusion sets for single use, gravity feed.
This document specifies requirements for single use, gravity feed infusion sets for medical use in order to ensure their compatibility with containers for infusion solutions and intravenous equipment.
Secondary aims of this document are to provide guidance on specifications relating to the quality and performance of materials used in infusion sets and to present designations for infusion set components.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.
ISO 3696, Water for analytical laboratory use — Specification and test methods
ISO 7864, Sterile hypodermic needles for single use — Requirements and test methods
ISO 8536-13, Infusion equipment for medical use — Part 13: Graduated flow regulators for single use with fluid contact
ISO 8536-14, Infusion equipment for medical use — Part 14: Clamps and flow regulators for transfusion and infusion equipment without fluid contact
150 14644-1, Cleanrooms and associated controlled environments — Part 1: ClassIfication of air cleanliness by particle concentration
ISO 15223-1, Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied — Part 1: General requirements
ISO 80369-7, Small-bore connectors for liquids and gases in healthcare applications — Part 7: Connectors for intravascular or hypodermic applications
3 Terms and definitions
No terms and dehnitions are listed in this docLiment.
5 Designation
Designation shall follow label requirements according to Clause 10.
6 Materials
The materials from which the infusion set, its components and the stand-alone air-inlet device are manufactured (as described in Clause 4) shall comply with the requirements specified in Clause 7. Where components of the infusion set come into contact with solutions, the materials shall also comply with the requirements specified in ClausesB and 9.
7 Physical requirements
7.1 Particulate contamination
The infusion set and stand-alone air-inlet device shall be manufactured under conditions that minimize particulate contamination. All parts shall he smooth and clean at the fluid pathway surfaces. When tested as specified in Al. the number of particles shall not exceed the contamination index limit.
7.2 Leakage
The infusion set, when tested in accordance with A.3. shall show no signs of air leakage.
7.3 Tensile strength
When tested as specified in A.4., the infusion set, excluding protective caps, shall withstand a static tensile force of not less than 15 N for 15 s.BS EN ISO 8536-4 pdf downlaod.


Tags:
Related Downloads
  • ISO 8600-1-2013 pdf free download.Endoscopes  – Medical endoscopes and endotherapy devices – Part 1: General requirements. 6 Marking 6.1 Minimum marking Iach individual endoscope or endotherapy device shall have the lollowing minimum marking: a) model number and/or other mark sufficient to identify the endoscope or endotherapy device and its manufacturer; b) maximum insertion portion width, minimum instrument channel width, working length, field of view and/or direction of view where such identification is necessary for the intended use of the endoscope or endotherapy device. The Insertion portion width and instrument channel width units shall be in millimetres. The insertion portion width and instrument channel width can also be marked in French size as defined in 3.5, shown by either ‘Fr or an encircled number; c) wherever reasonable and practicable. the endoscope or endotherapy device and detachable component(s) shall be identified in terms of lot numbers or serial numbers, etc. 6.2 Marking legibility The marking shall remain legible over the lifetime of the device when the endoscope or endotherapy device is used, cleaned, disinlected, sterilized and stored in accordance with the instruction manual. 6.3 Marking exceptions When marking on the endoscope or endotherapy device or detachable component(s) is impossible to achieve due…

  • ISO 15688-2012 pdf free download.Road construction and maintenance equipment – Soil stabilizers 一 Terminology and commercial specifications. This International Standard establishes the terminology, definitions of operation and commercial specifications for soil stabilizers and their components intended for use in road construction and pavement works. It does not apply to soil stabilizers used in agricultural applications. 2 Normative references The following documents, in whole or in part, are normatively referenced in this document and are indispensable for its application. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.. ISO 3911, Wheels and rims for pneumatic tyres — Vocabulary, designation and marking ISO 6746-1, Earth-moving machinery — Definitions of dimensions and codes — Part 1: Base machine ISO 6746-2, Earth-moving machinery — Definitions of dimensions and codes — Part 2: Equipment and attachments ISO 7134. Earth-moving machinery — Graders — Terminology and commercial specifications 3 Terms and definitions For the purposes of this document, the following terms and definitions apply. 3.1 soil stabilizer self-propelled machine, either towed or transported, with the function of pulverizing, breaking-up, aerating, homogenizing, and loosening existing and imported soil or paving materials and mixing them…

  • ISO 18436-1-2012 pdf free.Condition monitoring and diagnostics of machines – Requirements for qualification and assessment of personnel一 Part 1: Requirements for assessment bodies and the assessment process. 5 Requirements for assessment body personnel 5.1 General provisions In order to ensure that the assessment process is carried out effectively and uniformly, the competence requirements for personnel involved in the entire process shall be defined by the assessment body and, in the case of a third party, be approved by the responsible TCC (in accordance with ISO/IEC 17024). The assessment body shall require its personnel (internal or external) to sign a contract, or other document, by which they commit themselves to comply with the rules defined by the assessment body, including those relating to confidentiality and those relating to independence from commercial and other interests, and from any prior or present link with the persons to be examined that would, in the opinion of the interested parties, compromise impartiality. Clearly documented instructions shall be available to the personnel, describing their duties and responsibilities. These instructions shall be maintained up to date. All personnel involved in any aspect of assessment activities shall possess appropriate educational qualifications, experience and technical expertise, which satisfy defined…